UltraMem50 Crossflow Cassette
Online Ordering
| Specs | Number | Description | Price |
|---|
Related Documents
![]() |
| Tangential Flow UF Cassette User Manual |
Features
- UltraMem 50 is easy to use: plug, connect, ready to run
- Expandable modular design: each UltraMemTM 50 cassette is an independent module and easy to connect with other modules. The expansion of multi-modules can linearly scale up the amount of liquid to be processed, up to 3000 ml
- Innovative structural design: the specially designed inner flow path benefits the TFF process efficiency, lowers the shear force to fluid as well, friendly ro shear-sensitive liquids
- Multiple high performance MWCO PES ultrafiltration membrane for you to choose
- Safety: good integrity, low E&L and broadchemical compatiblity. Latest bio-pharm regulations USP 87 and 88 are met.
Applications
Purifications and concentration of:
- mAB, recombinant proteins
- Virus vectors and vaccines
- Blood-derivative products
- Plasmid, mRNA LNP
- Particulate dispersions
Technical Parameters
| Product Information |
Device type | TFF or Crossflow |
| L/H/W | 115×90×24 mm | |
| Effecive membrane area | 50cm² | |
| Materials of Construction | Clamp | Polypropylene |
| Housing material | Polycarbonate | |
| Seals & O-ring | Medical Silicone | |
| Membrane material | Polyether sulfone (PES) | |
| Operation | Maximum allowable operating Temp. | 60℃(140℉) |
| Maximum allowable(working)pressure | 3bar(44psi) | |
| Parameters | Available MWCO | 3/5/10/20/30/50/100 kDa |
| Volume can be Handled | 100-3,000 ml | |
| Min.Recirculating Volume | 20 ml |
FAQ
Q1: How to select the appropriate transmembrane pressure?
A: Transmembrane pressure depends on feed rate, solution viscosity, membrane pore size, and restrictor adjustment.
We recommend 1‒2 bar (max 3 bar). For first time use, start at 1 bar and increase gradually based on permeate flow.
Q2: What flow rate should be set for the peristaltic pump?
A: Recommended flow rate: 200‒400 mL/min. Higher flow rates provide higher transmembrane pressure and better
filtration efficiency.
Q3: How to perform a water flux test?
A: Set a suitable feed rate, adjust the restrictor to achieve 1 bar transmembrane pressure, measure the permeate
volume (L), then divide by membrane area (m²) and time (h) to obtain LMH.
Q4: How to maximize recovery of target molecules?
A: After concentration, rinse the cassette and tubing with a small volume of buffer, then combine the rinse with the
retentate. Typically >95% of the initial sample can be recovered.
Q5: How to select the correct membrane pore size?
A: For ultrafiltration concentration, choose an MWCO 3‒5 times smaller than the target protein (e.g., for a 150 kDa
protein, use 30 kDa or 50 kDa cassette). For clarification (e.g., E. coli/yeast lysate, CHO cell culture media), choose
0.22 μm or 0.45 μm.
Q6: Should I run a single cassette or multiple cassettes in series?
A: Use a single cassette for initial trials or sample volume ≤ 300 mL. For sample volumes up to 1 L or to increase
processing speed, connect cassettes in series.
Q7: Do I need extra accessories to connect two cassettes?
A: No. Each box already includes the necessary tubing for series connection.
Q8: Can the cassette be reused?
A: Yes, but no more than 3 times is recommended. Actual reuse number depends on the sample.
Q9: Can the pump tubing be reused?
A: Pump tubing wears out under peristaltic action. Replace every 2‒3 runs.
Q10: How is the chemical compatibility?
A: Stable in most common aqueous buffers and some solvents (e.g., 70% ethanol). See Section 3.4.
Q11: How to handle leakage?
A: Leakage occurs in two scenarios. Please refer to the following table based on the observed phenomenon.
Case 1 ‒ Target molecule appears in the permeate
| Possible Cause | Solution |
| Incorrect MWCO selection | Choose an MWCO 3-5 times smaller than the target. For example: Fora 150 kDa protein, a 30 kDa or 50 kDa cassette is recommended. |
| Chemical incompatibility damaging the membrane | Referto the chemical compatibility table in Section 4 to confirm whether the membrane material is compatible with the solution. |
| Over reused cassette, enlarged pores | It is recommended to reuse each cassette no more than 3 times, or replace it with a new cassette. |
Case 2 ‒ Visible liquid seepage from the cassette housing
| Possible Cause | Solution |
| Clogged permeate outlet causing pressure build up |
Carefully insert a syringe needle into the permeate outlet and gently clear the blockage (be careful not to use excessive force to avoid damaging the cassette) |
| Loose Luer connections | Use a clamp to appropriately tighten the connection. being careful not to over tighten as this may cause cracking. |
| Pump tubing fatigue leading to leakage at fittings |
It is recommended to replace the peristaltic pump tubing every 1-2 uses. as tubing under prolonged pressure will lose its elasticity. |